beautypg.com

Chapter 5, Error messages – B&K Precision MDL Series - Programming Manual User Manual

Page 93

background image

93

LIST:WIDth 1,10ms

7.

Specify the current setting for step 2.

LIST:LEVel 2,10

8.

Set the current slew rate for step 2.

LIST:SLEW 2,1

9.

Set the width for step 2.

LIST:WIDth 2,10ms

10.

Specify the current setting for step 3.

LIST:LEVel 3,20

11.

Set the current slew rate for step 3.

LIST:SLEW 3,1

12.

Set the width for step 3.

LIST:WIDth 3,10ms

13.

Specify the current setting for step 4.

LIST:LEVel 4,15

14.

Set the current slew rate for step 4.

LIST:SLEW 4,1

15.

Set the width for step 4.

LIST:WIDth 4,10ms

16.

Save the list to memory location 2.

LIST:SAV 2

17.

Set the input regulation mode to be

controlled by values in a list.

FUNCtion:MODE LIST

18.

Set trigger source to BUS

TRIG:SOUR BUS

19.

Run the list file

*TRG

Chapter 5

Error Messages

Error Number List

This appendix gives the error numbers and descriptions that are returned by the electronic load. Error

numbers are returned in two ways:

1)

Error numbers are displayed on the front panel.

2)

Error numbers and messages are read back with the SYSTem:ERRor? query. SYSTem:ERRor?

returns the error number into a variable and returns two parameters, an NR1 and a string.

The following table lists the errors that are associated with SCPI syntax errors and interface problems.

It also lists the device dependent errors. Information inside the brackets is not part of the standard

error message, but is included for clarification. When errors occur, the Standard Event Status register

records them in bit 2, 3, 4, or 5: