B&K Precision MDL Series - Programming Manual User Manual
Page 46
![background image](/manuals/320753/46/background.png)
46
Parameters 0 to 1000 | MINimum | MAXimum | DEFault
*RST Value 1000
Examples TRAC:POIN 10
Query Syntax TRACe: POINts? [ MINimum | MAXimum | DEFault ]
Returned Parameters
Related Commands TRAC:FEED
TRACe:FEED
Channel Specific
This command is used to select the source of readings to be placed in the buffer. With VOLTage
selected, voltage readings are placed in the buffer (TRAC:POIN maximum value is 1000). With CURRent
selected, current readings are placed in the buffer (TRAC:POIN maximum value is 1000). With TWO
selected, voltage and current are placed in the buffer when storage is performed (TRAC:POIN
maximum value is 500).
Command Syntax TRACe:FEED
Parameters VOLTage | CURRent | TWO
*RST Value TWO
Examples TRAC:FEED VOLT
Query Syntax TRACe:FEED?
Returned Parameters
Related Commands TRAC:POIN
TRACe:FEED:CONTrol
Channel Specific
This command is used to select the buffer control. With NEVer selected, storage into the buffer is
disabled. When NEXT is selected, the storage process starts, fills the buffer and then stops. The buffer
size is specified by the :POINts command.
Command Syntax TRACe:FEED:CONTrol
Parameters NEVer | NEXT
*RST Value NEVer
Examples TRAC:FEED:CONT NEXT
Query Syntax TRACe: FEED:CONT?
Returned Parameters
Related Commands TRAC:FEED
TRACe:DATA?