B&K Precision MDL Series - Programming Manual User Manual
Page 63
63
[SOURce:]RESistance:TRANsient:ALEVel
[SOURce:]RESistance:TRANsient:BLEVel
Channel Specific
This command specifies the transient level of the input resistance. The transient function switches
between level A and level B.
Command Syntax [SOURce:]RESistance:TRANsient:ALEVel
[SOURce:]RESistance:TRANsient:BLEVel
Parameters 0 to max. rated resistance.
Unit Ω (ohms)
Examples RES:TRAN:ALEV 5 RES:TRAN:BLEV 0.5
Query Syntax [SOURce:]RESistance:TRANsient:ALEVel?
[SOURce:]RESistance:TRANsient:BLEVel?
Returned Parameters
Related Commands RES
[SOURce:]RESistance:TRANsient:AWIDth
[SOURce:]RESistance:TRANsient:BWIDth
Channel Specific
This command specifies the transient pulse width of the input resistance.
Command Syntax [SOURce:]RESistance:TRANsient:AWIDth
[SOURce:]RESistance:TRANsient:BWIDth
Parameters 1 to 65535 us
Unit S (second)
*RST Value 500 uS
Examples RES:TRAN:AWID 0.001 RES:TRAN:BWID 0.02
Query Syntax [SOURce:]RESistance:TRANsient:AWIDth?
[SOURce:]RESistance:TRANsient:BWIDth?
Returned Parameters
Related Commands RES
[SOURce:]POWer
Channel Specific
This command sets the power of the load when operating in constant power mode.