[SOURce:]RESistance:TRANsient:ALEVel
[SOURce:]RESistance:TRANsient:BLEVel

Channel Specific
This command specifies the transient level of the input resistance. The transient function switches

between level A and level B.

Command Syntax [SOURce:]RESistance:TRANsient:ALEVel

[SOURce:]RESistance:TRANsient:BLEVel

Parameters 0 to max. rated resistance.

Unit Ω (ohms)

Examples RES:TRAN:ALEV 5 RES:TRAN:BLEV 0.5

Query Syntax [SOURce:]RESistance:TRANsient:ALEVel?

[SOURce:]RESistance:TRANsient:BLEVel?

Returned Parameters

Related Commands RES

[SOURce:]RESistance:TRANsient:AWIDth
[SOURce:]RESistance:TRANsient:BWIDth

Channel Specific
This command specifies the transient pulse width of the input resistance.

Command Syntax [SOURce:]RESistance:TRANsient:AWIDth

[SOURce:]RESistance:TRANsient:BWIDth

Parameters 1 to 65535 us

Unit S (second)

*RST Value 500 uS

Examples RES:TRAN:AWID 0.001 RES:TRAN:BWID 0.02

Query Syntax [SOURce:]RESistance:TRANsient:AWIDth?

[SOURce:]RESistance:TRANsient:BWIDth?

Returned Parameters

Related Commands RES


[SOURce:]POWer

Channel Specific
This command sets the power of the load when operating in constant power mode.