1.4 Programming the Status Registers

You can use status register programming to determine the operating condition of the electronic load at

any time. For example, you may program the electronic load to generate an interrupt (assert SRQ)

when an event such as a current protection occurs. When the interrupt occurs, your program can then

act on the event in the appropriate fashion.

The following table defines the status bits. Figure shows the status register structure of the electronic

load. The Standard Event, Status Byte, and Service Request Enable registers and the Output Queue

perform standard GPIB functions as defined in the IEEE 488.2 Standard Digital Interface for

Programmable Instrumentation. The Operation Status and Questionable Status registers implement

functions that are specific to the electronic load.

Bit Configurations of Status Registers

Operation Status Group

Bit

Signal Meaning

0

CAL

Calibrating. The electronic load is computing new calibration constants.

5

TRG

Waiting. The electronic load is waiting for a trigger.

Channel Status Group

Bit

Signal Meaning

0

VF

Voltage Fault. Either an overvoltage or a reverse voltage has occurred. This bit

reflects the active state of the FLT pin on the back of the unit. The bit remains

set until the condition is removed and INP:PROT:CLE is programmed.

1

OC

Overcurrent. An overcurrent condition has occurred. This occurs if the current

exceeds 102% of the rated current or if it exceeds the user-programmed

current protection level. Removing the overcurrent condition clears the bit. If

the condition persists beyond the user programmable delay time, PS bit is also

set and the input is turned off. Both bits remain set until the condition is

removed and INP:PROT:CLE is programmed.

2

RS

Remote Sense. When the remote sense is connected, this bit is set.