Examples CURR:PROT 2
Query Syntax [SOURce:]CURRent:PROTection:LEVel?
Returned Parameters
Related Commands CURR:PROT:DEL CURR:PROT:STAT
[SOURce:]CURRent:PROTection:DELay
Channel Specific
This command specifies the time that the input current can exceed the protection level before the
input is turned off.
Command Syntax [SOURce:]CURRent:PROTection[:DELay]
Parameters 0 to 60 seconds
Unit seconds
*RST Value 0
Examples CURR:PROT:DEL 5
Query Syntax [SOURce:]CURRent:PROTection:DELay?
Returned Parameters
Related Commands CURR:PROT CURR:PROT:STAT
[SOURce:]CURRent:TRANsient:MODE
Channel Specific
This command selects the operating mode of the transient generator as follows in constant current
mode.
CONTinuous The transient generator puts out a continuous pulse stream after receipt of a trigger.
PULSe
The transient generator puts out a single pulse upon receipt of a trigger.
TOGGle
The transient generator toggles between two levels upon receipt of a trigger.
Command Syntax [SOURce:]CURRent:TRANsient:MODE
Parameters CONTinuous | PULSe | TOGGle
*RST Value CONTinuous
Examples CURR:TRAN:MODE TOGG
Query Syntax [SOURce:]CURRent:TRANsient:MODE?
Returned Parameters
Related Commands CURR:TRAN:ALEV TRAN
[SOURce:]CURRent:TRANsient:ALEVel
[SOURce:]CURRent:TRANsient:BLEVel