The simplest SCPI command is a single message unit consisting of a command header (or keyword)
followed by a message terminator. The message unit may include a parameter after the header. The
parameter can be numeric or a string.
VOLTage 20
Headers
Headers, also referred to as keywords, are instructions recognized by the electronic load. Headers may
be either in the long-form or the short-form. In the long-form, the header is completely spelled out,
such as VOLTAGE, STATUS, and DELAY. In the short form, the header has only the first three or four
letters, such as VOLT, STAT, and DEL.
Query Indicator
Following a header with a question mark turns it into a query.
VOLTage?, CURRent:PROTection?
If a query contains a parameter, place the query indicator at the end of the last header.
CURRent:PROTection? MAX
Message Unit Separator
When two or more message units are combined into a compound message, separate the units with a
semicolon.
STATus:OPERation?;QUEStionable?
Root Specifier
When it precedes the first header of a message unit, the colon becomes the root specifier. It tells the
command parser that this is the root or the top node of the command tree.
Message Terminator
A terminator informs SCPI that it has reached the end of a message. Three permitted message
terminators are:
1.
Newline (
2.
End or identify (
3.
Both of the above (
In the examples of this guide, there is an assumed message terminator at the end of each message.