Query Syntax PROG:ACT:CHAN?
Returned Parameters
:PROGram:ACTive:SEQuence
The command is used to specify the sequence to enable for the program. The sequence number is
linked to the instrument settings saved in the internal storage memory. Each sequence number in
each programs are linked to one of the instrument settings storage locations.
Sequence 1 to 10 of program 1 is linked (in order) to storage locations 1-10.
Sequence 1 to 10 of program 2 is linked (in order) to storage locations 11-20.
….
Sequence 1 to 10 of program 10 is linked (in order) to storage locations 91-100.
Command Syntax PROGram:ACTive:SEQuence
Parameters 0 – 512
The value, in decimal, is based on the binary
representation of the sequence number which are
represented by the bits, according to the following table.
1 – Selects that sequence number.
0 – Not selected.
Bit Value in
decimal
512
256
128
64
32
16
8
4
2
1
Bit
9
8
7
6
5
4
3
2
1
0
Sequence
10
9
8
7
6
5
4
3
2
1
Examples PROG:ACT:SEQ 384
384 = 110000000
Enables sequence 8 and 9.
Query Syntax PROG:ACT:SEQ?
Returned Parameters
:PROGram:SEQuence:SHORt
The command is used to specify the short sequence for the program. When a sequence is selected for
short, all channels for that sequence will be shorted for testing.
Command Syntax PROGram:SEQuence:SHORt
Parameters
The value, in decimal, is based on the binary
representation of the channels to which are represented
by the bits, according to the following table. Note: