beautypg.com

B&K Precision MDL Series - Programming Manual User Manual

Page 43

background image

43

Examples STAT:OPER:COND?

Returned Parameters (register value)

Related Commands STAT:QUES:COND?

STATus:OPERation:ENABle

This command and its query can be used to set and read the value of the Operation Enable register.

This register is a mask for enabling specific bits from the Operation Event register to set the operation

summary bit (OPER) of the Status Byte register. The operation summary bit is the logical OR of all

enabled Operation Event register bits. This command is not channel specific and applies to the entire

mainframe.

Command Syntax STATus:OPERation:ENABle

Parameters 0 to 65535

Default Value 0

Examples STAT:OPER:ENAB 32 STAT:OPER:ENAB 1

Query Syntax STATus:OPERation:ENABle?

Returned Parameters (register value)

Related Commands STAT:OPER?

STATus:QUEStionable?

This query returns the value of the Questionable Event register. The Event register is a read-only

register that holds (latches) all events that pass into it. Reading the Questionable Event register clears

it. This command is not channel specific and applies to the entire mainframe.

Query Syntax STATus:QUEStionable[:EVENt]?

Parameters None

Examples STAT:QUES:EVEN?

Returned Parameters (register value)

STATus:QUEStionable:CONDition?

This query returns the value of the Questionable Condition register. This is a read-only register that

holds the real-time (unlatched) questionable status of the electronic load. This command is not

channel specific and applies to the entire mainframe.

Query Syntax STATus:QUEStionable:CONDition?