Command Syntax [SOURce:]CURRent:RANGe
Parameters 0 through max. Rated current
Unit A (amps)
*RST Value max. current (high range)
Examples SOUR:CURR:RANGE 5
Query Syntax [SOURce:]CURRent:RANGe
Returned Parameters
Related Commands CURR CURR:SLEW
[SOURce:]CURRent:SLEW
Channel Specific
This command sets the slew rate for all programmed changes in the input current level of the
electronic load. This command programs both positive and negative slew rates. MAXimum sets the
slew to the fastest possible rate. MINimum sets the slew to the slowest rate.
Command Syntax [SOURce:]CURRent:SLEW[:BOTH]
Parameters See specifications
Unit A (amps per micro second)
Examples CURR:SLEW MAX
Related Commands CURR CURR:NEG CURR:SLEW:POS
[SOURce:]CURRent:SLEW:POSitive
Channel Specific
This command sets the slew rate of the current for positive transitions. MAXimum sets the slew to the
fastest possible rate. MINimum sets the slew to the slowest rate.
Command Syntax [SOURce:]CURRent:SLEW:POSitive
Parameters See specifications
Unit A (amps per micro second)
Examples CURR:SLEW:POS 0.0001
Query Syntax [SOURce:]CURRent:SLEW:POSitive?
Returned Parameters