51

Channel Specific
This command turns the transient generator on or off.

Command Syntax [SOURce:]TRANsient[:STATe]

Parameters 0 | 1 | OFF | ON

*RST Value OFF

Examples TRAN 1

Query Syntax [SOURce:]TRANsient[:STATe]?

Returned Parameters 0 | 1

Related Commands CURR:TRAN:CURR:MODE CURR:TRAN:ALEV

[SOURce:]PROTection:CLEar

Channel Specific
This command clears the latch that disables the input when a protection condition such as overvoltage

(OV) or overcurrent (OC) is detected. All conditions that generated the fault must be removed before

the latch can be cleared. The input is then restored to the state it was in before the fault condition

occurred.

Command Syntax [SOURce:]INPut:PROTection:CLEar

Parameters None

Examples :PROT:CLE

[SOURce:]INPut:TIMer

Channel Specific
These commands enable or disable the electronic load on timer.

Command Syntax [SOURce:]INPut:TIMer[:STATe]

Parameters 0 | 1 | OFF | ON

*RST Value OFF

Examples INP:TIM 1

Query Syntax INPut:TIMer[:STATe]?

Returned Parameters 0 | 1

Related Commands INP:TIM:DEL

[SOURce:]INPut:TIMer:DELay

Channel Specific