B&K Precision MDL Series - Programming Manual User Manual
Page 51
51
Channel Specific
This command turns the transient generator on or off.
Command Syntax [SOURce:]TRANsient[:STATe]
Parameters 0 | 1 | OFF | ON
*RST Value OFF
Examples TRAN 1
Query Syntax [SOURce:]TRANsient[:STATe]?
Returned Parameters 0 | 1
Related Commands CURR:TRAN:CURR:MODE CURR:TRAN:ALEV
[SOURce:]PROTection:CLEar
Channel Specific
This command clears the latch that disables the input when a protection condition such as overvoltage
(OV) or overcurrent (OC) is detected. All conditions that generated the fault must be removed before
the latch can be cleared. The input is then restored to the state it was in before the fault condition
occurred.
Command Syntax [SOURce:]INPut:PROTection:CLEar
Parameters None
Examples :PROT:CLE
[SOURce:]INPut:TIMer
Channel Specific
These commands enable or disable the electronic load on timer.
Command Syntax [SOURce:]INPut:TIMer[:STATe]
Parameters 0 | 1 | OFF | ON
*RST Value OFF
Examples INP:TIM 1
Query Syntax INPut:TIMer[:STATe]?
Returned Parameters 0 | 1
Related Commands INP:TIM:DEL
[SOURce:]INPut:TIMer:DELay
Channel Specific