beautypg.com

B&K Precision MDL Series - Programming Manual User Manual

Page 76

background image

76

Channel 18-11 represents channels 8 to 1 for MDL002

extended mainframe.

1 – channel will be shorted.

0 – channel will not be shorted.

Bit

15 14 13 12 11 10 9

8

7

6

5

4

3

2

1

0

Channel 18 17 16 15 14 13 12 11 8

7

6

5

4

3

2

1

Examples PROG:SEQ:SHOR 683

683 = 1000100

Sets channel 3 and 7 for short testing.

Query Syntax PROG:SEQ:SHOR?

Returned Parameters

:PROGram:SEQuence:ONTime

The command is used to specify the input ON time for the selected sequence, which controls how long

to enable the inputs for the selected sequence.

Command Syntax PROGram:SEQuence:ONTime ,

Parameters - Sequence number, 1 to 10.

- On time, 0.01 to 60.0 seconds.

Examples PROG:SEQ:ONT 4, 5

Sets sequence 4 ON time to 5 seconds.

Query Syntax PROG:SEQ:ONT?

Returned Parameters

:PROGram:SEQuence:OFFTime

The command is used to specify the input OFF time for the selected sequence, which controls how long

to disable the inputs for the selected sequence.

Command Syntax PROGram:SEQuence:OFFTime ,

Parameters - Sequence number, 1 to 10.

- Off time, 0.01 to 60.0 seconds.

Examples PROG:SEQ:OFFT 4, 5

Sets sequence 4 OFF time to 5 seconds.

Query Syntax PROG:SEQ:OFFT?

Returned Parameters