beautypg.com

B&K Precision MDL Series - Programming Manual User Manual

Page 49

background image

49

[SOURce:]INPut:SYNCon

Channel Specific
This command enables or disables to change the electronic load inputs when INP:ALL command is

received.

Command Syntax [SOURce:]INPut:SYNCon[:STATe]

Parameters 0 | 1 | OFF | ON

*RST Value ON

Examples INP:SYNC 1

Query Syntax INPut:SYNCon[:STATe]?

Returned Parameters 0 | 1

Related Commands INP:ALL

[SOURce:]INPut:SHORt

Channel Specific
This command programs the specified electronic load module to the maximum current that it can sink

in the present operating range.

Command Syntax [SOURce:]INPut:SHORt[:STATe]

Parameters 0 | 1 | OFF | ON

*RST Value OFF

Examples INP:SHOR 1

Query Syntax INPut:SHORt:STATe?

Returned Parameters 0 | 1

Related Commands INP

[SOURce:]REMote:SENSe

Channel Specific
This command is used to select the remote measure mode for the load.

Command Syntax [SOURce:]REMote:SENSe[:STATe]

Parameters 0 | 1 | OFF | ON

*RST Value 0

Examples REM:SENS 0