Trigger Commands

The trigger subsystem is made up of a series of commands and subsystems to configure the trigger

model.

TRIGger:SOURce

This command selects the trigger source. This command is not channel specific and applies to the

entire mainframe.

BUS

Accepts a GPIB signal or a *TRG command as the trigger source. This

selection guarantees that all previous commands are complete before the trigger

occurs.

EXTernal

Selects the electronic load’s trigger input as the trigger source. This trigger is

processed as soon as it is received.

HOLD

Only the TRIG:IMM command will generate a trigger in HOLD mode. All other

trigger commands are ignored.

MANUal

The event occurs when the Trig key is pressed.

TIMer

This generates triggers that are in synchronization with the electronic load's

internal oscillator as the trigger source. The internal oscillator begins running as

soon as this command is executed. Use TRIG:TIM to program the oscillator period.

Command Syntax TRIGger:SOURce

Parameters BUS | EXTernal | HOLD | MANUal | TIMer

*RST Value MANUal

Examples TRIG:SOUR BUS TRIG:SOUR EXT

Query Syntax TRIGger:SOURce?

Returned Parameters

Related Commands ABOR TRIG TRIG:DEL

TRIGger:TIMer

This command specifies the period of the triggers generated by the internal trigger generator. This

command is not channel specific and applies to the entire mainframe.

Command Syntax TRIGger:TIMer

Parameters 1 to 999.99s | MINimum | MAXimum | DEFault

Unit seconds