beautypg.com

B&K Precision MDL Series - Programming Manual User Manual

Page 65

background image

65

[SOURce:]POWer:PROTection

Channel Specific
This command sets the soft power protection level. If the input power exceeds the soft power

protection level for the time specified by POW:PROT:DEL, the input is turned off.

NOTE: Use POW:PROT:DEL to prevent momentary power limit conditions caused by programmed

changes from tripping the overpower protection.

Command Syntax [SOURce:]POWer:PROTection[:LEVel]

Parameters 0 to max. rated power

Unit W (power)

*RST Value MAXimum

Examples POW:PROT 100

Query Syntax [SOURce:]POWer:PROTection[:LEVel]?

Returned Parameters

Related Commands POW:PROT:DEL

[SOURce:]POWer:PROTection:DELay

Channel Specific
This command specifies the time that the input power can exceed the protection level before the input

is turned off.

Command Syntax [SOURce:]POWer:PROTection:DELay]

Parameters 0 to 60 seconds

Unit seconds

*RST Value 0

Examples POW:PROT:DEL 5

Query Syntax [SOURce:]POWer:PROTection:DELay?

Returned Parameters

Related Commands POW:PROT

[SOURce:]POWer:CONFig

Channel Specific
This command sets the hard power protection level.