Source commands – B&K Precision MDL Series - Programming Manual User Manual
Page 48
48
Command Syntax TRACe:TIMer
Parameters 0.0002 to 3600s | MINimum | MAXimum | DEFault
Unit S (second)
*RST Value 1
Examples TRAC:TIM 0.1
Query Syntax TRACe:TIMer? [MINimum | MAXimum | DEFault]
Returned Parameters
Source Commands
These commands control the input of the electronic load. The INPut and OUTput commands are
equivalent. The CURRent, RESistance, and VOLTage commands program the actual input current,
resistance, and voltage.
[SOURce:]INPut:ALL
This command enables or disables all the inputs of the module. The state of a disabled input is a high
impedance condition.
Command Syntax [SOURce:]INPut:ALL[:STATe]
Parameters 0 | 1 | OFF | ON
Examples INP:ALL 1
[SOURce:]INPut
Channel Specific
This command enables or disables the electronic load inputs. The state of a disabled input is a high
impedance condition.
Command Syntax [SOURce:]INPut[:STATe]
Parameters 0 | 1 | OFF | ON
*RST Value OFF
Examples INP 1
Query Syntax INPut[:STATe]?
Returned Parameters 0 | 1
Related Commands *RCL *SAV