B&K Precision MDL Series - Programming Manual User Manual
Page 66
66
Command Syntax [SOURce:]POWer:CONFig[:LEVel]
Parameters 0 through max. Rated power
Unit W (power)
Examples POW:CONFig 100
Query Syntax [SOURce:]POWer:CONFig[:LEVel
Returned Parameters
Related Commands POW:PROT
[SOURce:]IMPedance
Channel Specific
This command sets the resistance value when in CZ mode.
Command Syntax [SOURce:]IMPedance[:LEVel][:IMMediate]
Parameters 0 to 7500 ohms.
Unit R (ohms)
*RST Value MAXimum
Examples IMP 5 IMP:LEV 3.5
Query Syntax [SOURce:]IMPedance[:LEVel][:IMMediate]?
Returned Parameters
Related Commands IMP:RANG
[SOURce:]IMPedance:RANGe
Channel Specific
This command is used to set the resistance range when the load is in CZ mode. This limits the settable
resistance.
Command Syntax [SOURce:]IMPedance:RANGe
Parameters 0 to 7500 ohms.
Unit R (ohms)
Examples IMP:RANG 15 SOUR:IMP:RANGE MIN
Query Syntax [SOURce:]IMPedance:RANGe?
Returned Parameters
[SOURce:]IMPedance:RESistance
Channel Specific