B&K Precision MDL Series - Programming Manual User Manual
Page 32
32
NOTE: The device state stored in location 0 is automatically recalled at power turn-on.
Command Syntax *RCL
Parameters 0 to 9
Example *RCL3
Related Commands *PSC *RST *SAV
*RST
This command resets ALL channels of the electronic load to the following factory-defined states:
CURR MIN
RES:TRAN:BLEV MIN
CURR:MODE FIX
RES:TRAN:BWID MIN
CURR:PROT:DEL 3
RES:TRAN:MODE CONT
CURR:PROT:LEV MAX
SENS:AVER:COUN 8
CURR:PROT:STAT OFF
SENS:AVER:STAT 1
CURR:RANG MAX
SENS:FUNC:CURR DC
CURR:SLEW MAX
SENS:NPLC 7
CURR:TRAN:ALEV MAX
SENS:VOLT:RANG:AUTO ON
CURR:TRAN:AWID MIN
TRAC:FEED TWO
CURR:TRAN:BLEV MIN
TRAC:FEED:MODE NEV
CURR:TRAN:BWID MIN
TRAC:POIN 2
CURR:TRAN:MODE CONT
TRAN OFF
FUNC CURR
TRIG:COUN 1
FUNC:MODE FIX
TRIG:DEL 0
INP OFF
VOLT MAX
INP:SHOR OFF
VOLT:ON MIN
POW:CONF MAX
VOLT:ON:LATC ON
POW:PROT:DEL 3
VOLT:RANG MAX
RES MAX
VOLT:TRAN:ALEV MAX
RES:RANG MAX
VOLT:TRAN:AWID MIN
RES:TRAN:ALEV MAX
VOLT:TRAN:BWID MIN
RES:TRAN:AWID MIN
VOLT:TRAN:MODE CONT