beautypg.com

Xylem CHATTER RTU and E-mail User Manual

Page 176

background image

USERS MANUAL

CHATTER™ DATA LOGGER

176

Change the output unit from bar to inches of water column

Output value (401,4742 1 Sensor value) (0 0)

Output value Sensor value 401,4742

Change the output unit from °Celsius to °Fahrenheit

Output value (1 1,8 Sensor value) (32 0)

Output value Sensor value 1,8 32

Change the output from level above sensor to depth to water surface

The standard setting for the Chatter™ Data Logger is to
measure the water level above the pressure transmitter, A in
the drawing to the right

If you need to get the depth to water instead, B in the
drawing to the right, you will have to change the settings.

You must know the sum of A and B, it equals the depth of the
sensor tip measured from the soil surface.

In the examples below, A is set to 3 meters and B is set to 5
meters, the sum is 8 meters.

The depth to water surface is always a negative
number!

Output value unit is in bar