Pvc tests – Paradyne CSU User Manual
Page 286
8. Troubleshooting
8-22
September
2002
9128-A2-GB20-80
PVC Tests
PVC tests can be run on a requested DLCI for a selected interface.
T
When PVC tests are on a multiplexed DLCI between FrameSaver devices,
they are nondisruptive to data, so user data can continue to be sent during a
test.
T
If the device at one end of the circuit is not a FrameSaver device, PVC tests
are on a standard DLCI and are disruptive to data. Also, the Connectivity test
would not appear.
Loopback, and send/monitor pattern tests are available for each interface on the
selected DLCI. FrameSaver devices should be at each end of the circuit. If a PVC
Loopback is started at one end of the circuit, the other end can send and monitor
pattern tests.
The example below shows a PVC Test screen for a FrameSaver unit with ISDN
backup capability, with the multiplexed DLCI 550 selected. If a standard DLCI was
selected, (
Disruptive
), rather than (
Non-Disruptive
), would be displayed
after Test. Also, the Connectivity test would not appear.
PVC Tests Screen Example
If the unit does not have the ISDN DBM feature, or if the ISDN Link Profile’s Link
Status is disabled,
Test Call
does not appear. An Outbound Phone Number
must be configured for
Test Call
to appear.
NOTE:
Errors encountered during these tests may be caused by mismatched CIRs in
the two FrameSaver units. If errors are detected, verify the CIR configuration
and retest.
main/test/network_pvc 9128-II
Device Name: Node A 5/26/2000 23:32
NETWORK PVC TESTS
DLCI Number: 550
Test (Non-Disruptive) Command Status Result
PVC Loopback: Start
Inactive 0:00:00
Send Pattern: Start
Inactive 0:00:00
Monitor Pattern: Start
Inactive 0:00:00
Sequence Errors 99999+
Data Errors 99999+
Connectivity: Start
Inactive RndTrip Time (ms)
99999
Test Call: Stop
Active Frame Relay Link Up
--------------------------------------------------------------------------------
Ctrl-a to access these functions, ESC for previous menu MainMenu Exit