beautypg.com

Optical edge detection is not working properly – HEIDENHAIN ND 1300 OED and Crosshair Systems User Manual

Page 241

background image

223

11

Problem Solving

Optical edge detection is not working properly

Probable cause 1: Reference or sensor light levels are incorrect

Possible solution:

1 Refer to Evaluating optical light levels in Chapter 10: Setup for instructions regarding

light levels, and then reorient, repair or replace light sources or optical cables.

Probable cause 2: Edge detection is not calibrated:

Possible solution:

1 Perform a Teach operation or a Distance Cal using the Magnifications Setup screen.

Probable cause 3: Edge detector and crosshair probes are not

cross-calibrated:

Possible solution:

1 Perform a Cross Cal using the Magnifications Setup screen.

Probable cause 4: Multiple system magnifications are not

calibrated to be parcentric:

Possible solution:

1 Perform a Parcentric calibration using the Magnifications Setup screen.

Symptoms, Probable Causes & Solutions