Super Systems 20PQ User Manual
Page 146

142
Temperature drift: (CJ excluded)
< 200 ppm/°C of span for mV and TC ranges 3, 4, 7, 18, 19, 22.
< 300 ppm/°C of span for mA/V
< 250 ppm/°C of span for TC range 1, 2, 5, 6, 8, 9, 10, 14, 16,
17, 20, 21, 23, 24, 25, 29.
< 350 ppm/°C of span TC ranges 11, 12, 13, 26, 27, 28.
< 500 ppm/°C of span for RTD.
Operative temperature: from 0 to 50 °C (+32 to 122 °F).
Storage temperature: -20 to +70 °C (-4 to 158 °F)
Humidity: from 20 % to 85% RH, non condensing.
INPUTS
A) THERMOCOUPLE
Type: J, K, T, E, N, S, R, B, L, U, G, D, C, Ni/Ni 18% Mo. °C/°F
selectable.
External resistance: 100
W max, maximum error 0,1% of span.
Cold junction: automatic compensation from 0 to 50 °C.
Cold junction accuracy : 0.1 °C/°C
Input impedance: > 1 M
W
Calibration : according to IEC 584-1 and DIN 43710 - 1977.
T/C
type
L
J
K
T
U
E
N
S
R
B
G
D
C
Ni
L
J
K
T
U
E
Ranges
1
2
3
4
5
6
7
8
9
10
11
12
13
14
16
17
18
19
20
21
-100 / 900
°C
-100 / 1000
°C
-100 / 1370
°C
-200 / 400
°C
-200 / 600
°C
-100 / 800
°C
-100 / 1400
°C
- 50 / 1760
°C
- 50 / 1760
°C
0 / 1820
°C
0 / 2300
°C
0 / 2300
°C
0 / 2300
°C
0 / 1100
°C
-150 / 1650 °F
-150 / 1830 °F
-150 / 2500 °F
-330 / 750 °F
-330 / 1110 °F
-150 / 1470 °F
DIN 43710-1977
IEC 584-1
IEC 584-1
IEC 584-1
DIN 43710-1977
IEC 584-1
IEC 584-1
IEC 584-1
IEC 584-1
IEC 584-1
DIN 43710-1977
IEC 584-1
IEC 584-1
IEC 584-1
DIN 43710-1977
IEC 584-1
XKP-1-C2.p65
10/30/01, 9:37 AM
142