Super Systems X5 Calibration Manual User Manual
Super Systems Equipment
X5 Controller
Calibration Guide
Super Systems Inc. X5 Controller: Calibration Guide. (Document #4615; New: 05/01/2014).
Introduction
The instrument calibration is based on a series of steps. Each step performs the calibration of a particular input or
output. It must be done only when strictly necessary and using reference instruments with a high accuracy. Then the
instrument must remain switched on at current environmental conditions for at least one hour.
Calibration Procedure
Starting from operation mode, press and hold both the “
↵
“(enter) and “
“(menu access) keys. “C.PaS” will
appear. Enter the Configuration Password (default: 33).
NOTE: See the X5 Controller product manual section “4.6 PARAMETERIZATION - ACCESS MENU - PASSWORD
- CALIBRATION” for more details on the Configuration Password.
“C.tc” (T/C calibration) will appear on the lower display while a value will appear on the upper display.
Generate a signal using the below table; use copper wire to source the signal. Press the “
∨
“ or “
∧
“ keys to
adjust the value to be as close to zero as possible: 0 ±2.
Step
Function
Action
C.tc.
TC calibration
Apply 50.000mV to the input – set with
∨
or
∧
keys
C.Pt.
RTD calibration
Apply 313.594 Ω to the input – set with
∨
or
∧
keys
C.mV.
mV calibration
Apply 300 mV to the input – set with
∨
or
∧
keys
C.Vol
Volts Calibration
Apply 10 V to the input – set with
∨
or
∧
keys
C.Ma
mA calibration
Apply 20mA to the input – set with
∨
or
∧
keys
C.S. V
Remote Setpoint Volts
calibration
Apply 10 V to the input – set with
∨
or
∧
keys
C.S. mA
Remote Setpoint mA
calibration
Apply 20mA to the input – set with
∨
or
∧
keys
Table 1 - Calibration types and actions
Wait until the displayed value remains stable: 0 ± (2 to 5).
IMPORTANT: Press the “↵” key to bypass a calibration step. DO NOT press the “
∨
“ or “
∧
“keys unless you
want to Calibrate that specific step.
When the calibration is correct, press the “↵” key to go to the next step. It is possible to not modify the
calibration. When the step has been displayed on screen DO NOT press the “
∨
“ or “
∧
“ keys; only press the
“↵” key until out of calibration mode.
–OR-
After “C.tc.”, press the “↵” key to re-enter normal operation.