Super Systems 20PQ User Manual
Page 118
114
Gr. P
Run time group P [r.Axx]
Program editing
This group is available when the instrument operates as
programmer only.
- Program selection - [r.A01]
Lower display: SE.xx if the selected program is a simple
program.
P.L.xx if the selected program is a linked
program.
Where xx is the number of segments of the
simple program or the number of simple
programs used for the linked program.
Ranges: - from 1 to 90 for simple program; after 90
- from L1 to L9 for linked program
NOTES:
1) When the selected program is a new program, the PRG
indicator will flash and xx will be equal to 0.
2) If no segment is available, the instrument will show the
existing programs only.
3) It is not possible to select a linked program (from L1 to L9) if
no simple program is programmed.
- Soak time engineering units selection
- [r.A02]
This parameter can be modified only when the selected program
is not running and it is not a linked program.
Upper display: Pr. xx
where xx is the number of the selected
program
Range: "hh.ññ" for hours and minutes
"ññ.ss" for minutes and seconds.
- Ramp time engineering units selection
- [r.A03]
This parameter can be modified only when the selected program
is not running and it is not a linked program.
Upper display: Pr. xx
where xx is the number of the selected
program
Range: "hh.ññ" for hours and minutes
"ññ.ss" for minutes and seconds.
"GrAd" for gradient in digit/hour.