beautypg.com

Measurement Computing eZ-PostView rev.2.0 User Manual

Page 66

background image

3-8 Menus

979595

eZ-Analyst

Calibrator
Type

This column is used to select Peak or RMS* as the calibrator type. Highlight a
cell, or range of cells, in the grid, then select Peak or RMS from the popup

menu.

Peak Uses the peak amplitude of the spectrum around the specified frequency of the

calibration signal.

RMS Uses a compensated overall level calculation to determine the RMS level of the

calibration signal, as specified in the Units field.

*RMS – Root-mean-square, is the square root of the arithmetic mean of the

squares of a set of numbers.

Calibrator
Scale

This column is used to select the calibrator scale to linear or to decibel.

Lin Select Lin if the calibrator is in linear scale.

dB Select dB if the calibrator scale is dB.

Note: While there are several definitions to dB, in our application we are

using dB to express the ratio of the magnitudes of two quantities equal
to 20 times the common logarithm of the ratio.

The formula for dB is:

dB = 20 log (x/dBEUref)

EU dB
Reference

This field is used to enter the Engineering Units dB reference.

dB EU Reference is applied to displayed data when the Y-axis scale is set to dB.

This is valid only for frequency domain data.
Note: The dB Reference (Volt) can be changed simultaneously for all channels

from the associated entry box in the FFT Setup Tab. The tab is accessed

from the Configuration Window [via the Edit pull-down menu].

The formula for dB display for Unaveraged Spectrum, Averaged Spectrum, and

FRF is:

dB = 20 log (x/dBEUref)

The formula for dB display for Autospectrum, Cross Spectrum, and PSD is:

dB =10 log [x/((dBEUref)^2)]


Calibrator
EU

This field is used to set the Calibrator’s Engineering Units.

Measured
mV

Displays the measured value of the channel in milli-volts.

Transducer
mV/EU

The Transducer mV/EU value can be entered manually, or by the
measurement process. Highlight a cell, or cells, in the grid before typing a
value in the data entry box. Then press the button to accept the

value. All other fields linked to this value are updated when the value is

accepted.

mV/EU= Volts/Units * Transducer Gain

EU Offset

EU Offset is used for DC signal compensation. Offset is added to, or
subtracted from the measured EU value.