beautypg.com

Calibration, Calibration……… -6, Task menu > calibration – Measurement Computing eZ-PostView rev.2.0 User Manual

Page 64

background image

3-6 Menus

979595

eZ-Analyst

Task Menu >

Calibration

When calibration is performed, a signal of known Peak level [or RMS value] is supplied to a

transducer that is connected to an active input channel. An accelerometer calibrator or piston

phone is typically used to generate the calibration signal for vibration sensors and microphones,
respectively.

Examples:

Accelerometer calibrators typically make use of linear engineering
units and, as their name implies, are used for calibrating

accelerometers.

Piston phones are most often used for calibrating microphones. Piston

phones typically make use of decibel (dB) engineering units.

eZ-Analyst includes a Calibration window for selecting the channels to be calibrated and for
entering several signal-related parameters. In addition, the calibration is actually started from

the window.

When you are in “Measurement Mode” you can access the Calibration window from the Task
Menu or from the Input/Output Channels tab.

Calibration Display Screen

¾ When a channel is calibrated, the number of averages used will be 5, or

the number that is designated in the “No. of Averages” field (located on

the Analyzer Tab). The greater of the two values will be used,

automatically.

¾ When a channel is calibrated, the blocksize of the acquisition will

automatically be 4096, or the blocksize value that is stated on the

Analyzer Tab, providing it is 4096

or higher.

A discussion of the various regions of the Calibration window now follows.

The section concludes with an example.