beautypg.com

Considerations regarding double hammer rejection, Considerations regarding double, Hammer rejection – Measurement Computing eZ-PostView rev.2.0 User Manual

Page 101

background image

eZ-Analyst

969795

Edit Menu 4-23

Edit Menu > Configuration >

Block Rejection Tab:

Hammer Rejection Panel

This panel is used to define the acceptable range for a hammer signal. Any signal outside of
the defined range is detected as a double hammer. The acceptable range will appear as a

rectangle whose boundaries are defined by high and low limits for the blocksize, and by high

and low limits for scale. Respectively, these are x- and y- axis limits.

X-axis Limits (% of blocksize): Used to set a Low point and a High point along the x-axis,
for which the hammer impact must occur to be acceptable. An example setting is 5% to 15%

(of blocksize).

Y-axis Limits (% of F(ull) S(cale): Used to set a Low point and a High point along the y-
axis, for which the hammer impact must occur to be acceptable. An example is –5.000 to

+ 5.000 (% of Full Scale).

Considerations Regarding Double Hammer Rejection
There are three major interrelated components to consider when setting up a double-hammer

rejection condition. These are:

Trigger Mode and Delay

Force Window

Double Hammer Region

A trigger is used to ensure time synchronous measurements across all the active channels.

As indicated in the

Trigger Panel

section of this document (beginning on page 4-6), a pre-

trigger indicates that data captured before a trigger event [the amount of which is specified

by the delay] will be prefixed to the data following the trigger event. Also, a trigger delay

[in reference to the Start Trigger Condition] indicates that a specified amount of data after

the start trigger [the amount of which is specified by the delay] will be skipped (ignored)
before a block of data is captured.

Double hammer reject is only meaningful with the pre-trigger specified

. Double

hammer rejection protects the measurement from a bad hammering.

It is desirable to set the pre-trigger to fall within the hammer region, since it is a common

practice to set the trigger channel to an impact channel.

Example
In this example, a test engineer is looking for a trigger event that occurs somewhere after

10% of the data block, but before 15% of the block. Therefore:

(1) Pre-Trigger is set for 10% Delay, using the

Trigger Panel

(on the Analyzer Tab).

(2) A Rectangular Force Window is selected and set for a Start point of 9% and a Stop

point of 15%. This is done in the FFT Setup Tab’s

FFT Window Functions Panel

. Page

4-27 contains additional information.

Selecting a Rectangular Force

Window and Setting Points

for 9% and 15%

Selected and set in the FFT Setup Tab.

A Double Hammer Event

When impact testing we want to make sure that a double hammer does not occur. Through

trial and error we may start by setting the Double Hammer range at 12% and 14%, or 11%

and 13% of the block. Typically this range is fine tuned using sample acquisitions prior to
starting a real test.