Fluke Biomedical medSim 300B User Manual
Page 92

medSim 300B
Operators Manual
3-64
Respiration Shift:
•
SFTOFF
Baseline shift, off
•
R+SFT
Baseline shift, positive
•
R-SFT
Baseline shift, negative
Blood Pressure—All Channels:
•
ZALL
Zero all blood pressure channels
Blood Pressure Channel 1:
•
P1ART Arterial
•
P1LV Left
ventricle
•
P1CVP
Central venous pressure
•
P1RV Right
ventricle
•
P1PA Pulmonary
artery
•
P1W
Pulmonary artery wedge
•
P1Sn
Static = n
n = -10, -5, 0, 20, 40, 80, 100, 200, 250, and 300
Blood Pressure Channel 2:
•
P2ART Arterial
•
P2LV Left
ventricle
•
P2CVP
Central venous pressure
•
P2RV Right
ventricle
•
P2PA Pulmonary
artery
•
P2W
Pulmonary artery wedge
•
P2Sn
Static = n
n = -10, -5, 0, 20, 40, 80, 100, 200, 250, and 300
Blood Pressure Channel 3:
•
P3CVP
Central venous pressure
•
P3RV Right
ventricle
•
P3PA Pulmonary
artery
•
P3W
Pulmonary artery wedge
•
P3Sn
static = n
n = -10, -5, 0, 5, 10, 20, and 30
•
STSG Swan-Ganz,
start
•
INS Swan-Ganz,
insert
•
INF
Swan-Ganz, inflate
•
DEF Swan-Ganz,
deflate
•
REM Swan-Ganz,
remove