2. Select the desired apnea mode by pressing the corresponding function key.
Apnea mode selections and their descriptions follow:
Selections Descriptions
off
Turns off apnea. Resumes breathing.
mom
Momentary. Causes apnea while the function key is pressed.
cont
Continuous apnea. Disabled by the "off" selection.
12 s
Generates 12 seconds of apnea each time the key is pressed.
32 s
Generates 32 seconds of apnea each time the key is pressed.
For respiration baseline shift:
1. Press
are available as shown:
resp base shift
off
pos
neg
eic067.eps
2. A selection of "positive" or "negative" shifts the respiration baseline either
positive or negative and changes the respiration rate to 120 BrPM for 12 of every
60 seconds.
Baseline shift selections and their descriptions follow:
Selections Descriptions
off
Turns off baseline shift.
pos
Shifts baseline up and decreases delta impedance to 1/6 for 12
seconds.
neg
Shifts baseline down and decreases delta impedance to 1/6 for 12
seconds.
Function Select - Blood Pressure Menu
To choose the BLOOD PRESSURE MENU, press
FUNCTION SELECT MENU.
The BP CHANNEL SELECT MENU offers the following five selections—1, 2, 3, 4 and
zero—on two pages of display:
BP CHANNEL SELECT
1
2
3
4
>
eic068.eps