Late deflation, Normal sinus rhythm simulation – Fluke Biomedical medSim 300B User Manual
Page 131

Option 1
Intra-Aortic Balloon Pump (IABP) Augmentation
4
4-33
Late Deflation
Inflation
Deflation
Dicrotic Notch
Assist Ratio = 1:2
15% Decrease
Assisted
Waveform
Unassisted
Waveform
eic209.eps
Figure 4-29. medSim 300B Simulated Late Balloon Deflation
•
Inflation begins at the dicrotic notch (aortic valve closure).
•
Late deflation causes a 15 % decrease in the following systolic peak pressure.
Normal Sinus Rhythm Simulation
For simulated normal sinus rhythm (NSR) waveforms:
1. Select
BP CHANNEL SELECT
1
2
3
>
4
eic210.eps
2. Press
BP 4
WAVE
STAT
IABP
eic211.eps
3. Press
INTRA AORTIC BALLOON PUMP
NSR
ARRH
PACE
eic212.eps
See also other documents in the category Fluke Biomedical Equipment:
- 7000DP Impulse Supplement (12 pages)
- 7000DP Impulse Getting Started (36 pages)
- 7000DP Impulse (92 pages)
- 06-526-2200 (26 pages)
- 07-417 (16 pages)
- 07-443 (14 pages)
- 07-451 (8 pages)
- 07-487 (12 pages)
- 07-453 (24 pages)
- 07-555 (8 pages)
- 07-553 (10 pages)
- 07-605-7777 (8 pages)
- 07-444 (46 pages)
- 07-618 (10 pages)
- 07-600 (6 pages)
- 07-591 (8 pages)
- SigmaPace 1000 (154 pages)
- 07-653 (8 pages)
- 07-633 (20 pages)
- 07-649 (14 pages)
- 07-661-7662 (12 pages)
- 07-645 (14 pages)
- 10100AT (80 pages)
- 07-644 (10 pages)
- 18-203 (4 pages)
- 07-621 (12 pages)
- 07-647 (12 pages)
- 18-207 (8 pages)
- 18-216-1000 (10 pages)
- 18-220 (10 pages)
- 18-228 (8 pages)
- 18-223 (18 pages)
- 18-229-1313 (6 pages)
- 18-250 (8 pages)
- 18-252 (6 pages)
- 18-303 (6 pages)
- 35035 (19 pages)
- 6000-528 (18 pages)
- 35080B (70 pages)
- 6000-529 (16 pages)
- 35080M (64 pages)
- 6000-530B (16 pages)
- 57-436 (12 pages)
- 57-440 (12 pages)
- 57-402 (32 pages)