Insertion selections, Pacemaker selections – Fluke Biomedical medSim 300B User Manual
Page 18

medSim 300B
Operators Manual
1-6
Ventricular Rhythm 2
Ventricular
Ventricular
tachycardia
Ventricular
fibrillation
Electromotive
disassociation
Asystole
Conduction
First degree A-V block
Second degree A-V block, type 1
Second degree A-V block, type 2
Third degree A-V block
Right bundle branch block
Left bundle branch block
Premature Ventricular Contraction (PVC) Parameter Selections
PVC Type
1, 2, 3 and 4
Note
See the "Operation" chapter for descriptions.
PVC Timing
R on T, early and standard
Note
See the "Operation" chapter for descriptions.
PVC per Minute
0 to 25. Manual and auto-stepping modes. Auto-step time
interval is 60 seconds.
PVC Total
Counts number of PVCs generated. Maximum count is 65535.
Insertion Selections
PVCs, PACs, and PNCs can be inserted one at a time by pushing a button. Operates on
normal sinus waveforms only and at ECG rates from 30 to 160 BPM.
Pacemaker Selections
Pacemaker Waveforms
Asynchronous at 75 BPM, continuously paced
Demand
1
Demand
2
Atrial-ventricular
sequential
Non-capture
Non-function
Atrial and ventricular pacemaker pulses are set up
independently