medSim 300B

Operators Manual

3-12

3. To set the amplitude for all ECG baseline waveforms, press -AMP from

page 1 of the BASELINE ECG MENU:

ecg amp, lead II: #.## mv

down

up

exc

auto

eic016.eps

This is a "step" menu. Refer to Step Menus in the MENU OVERVIEW section

earlier in this chapter for detailed information on automatic and manual step

functions.

4. To adjust the ST segment deviation (elevation and depression), press -ST

from page 1 of the BASELINE ECG MENU:

st deviation:

#.##

mv

down

up

exc

auto

eic017.eps

This function simulates an adult heart (not neonatal) at rates less than or equal to

160 BPM and with most arrhythmias having an 80 BPM rate.
This is a "step" menu. Refer to Step Menus in the MENU OVERVIEW section

earlier in this chapter for detailed information on automatic and manual step

functions.

5. Press -AXIS (from page 1 of the BASELINE ECG MENU) to set the axis

of the normal ECG to one of these three heart positions: intermediate, horizontal,

or vertical. The display appears as shown here:

axis deviation

int

horz

vert

eic018.eps

6. To select "neonatal ECG", press -NEO from page 2 of the BASELINE ECG

MENU. The display appears as illustrated:

neonatal (bp -> 75%)

off

on

eic019.eps