Operation
Operating the medSim 300B
3
3-17
premature arrhythmias
mf1
mf2
mf3
<
eic034.eps
The premature arrhythmia waveform selections can be made by pressing the
corresponding function key.
Descriptions of the premature arrhythmia selections follow:
Selections Descriptions
nsr
Normal sinus rhythm. Provides a convenient way to return to normal
sinus rhythm.
pac
Premature atrial contraction. Normal sinus rhythm except every 10
th
beat is premature by 25 %.
pnc
Premature nodal contraction. Nodal rhythm except every 10
th
beat is
premature by 25 %.
pvc
Premature ventricular contraction. Normal sinus rhythm except every
10
th
beat is a PVC.
mf1
Multifocal PVCs. Normal sinus rhythm except 2 of every 15 beats are
PVCs. PVC type alternates between type 1 and type 2.
mf2
Multifocal PVCs. Like "mf1" except PVC types are 2 and 3.
mf3
Multifocal PVCs. Like "mf1" except PVC types are 3 and 4.
The third group of arrhythmia waveforms is the ventricular rhythm 1 group—a group
with normal sinus rhythm at 80 BPM mixed with PVCs.
To select this group, press
following two display pages of six ventricular arrhythmia 1 waveforms:
ventricular arrh. 1
nsr
big
trig
pair
>
eic035.eps
ventricular arrh. 1
run5
run11
<
eic036.eps