Ecg performance waveform selections, Arrhythmia selections – Fluke Biomedical medSim 300B User Manual
Page 17

Chapter Title
medSim 300B Instrument Specifications
1
1-5
ECG Performance Waveform Selections
Amplitudes are set by "Baseline ECG Amplitude". Amplitude applies to Lead II and
V-leads. Lead I is 0.7 times Lead II; Lead III is 0.3 times Lead II. The V-leads are the
same as Lead II.
Performance Waveforms
Rates
Zero Zero
output
Pulse 4
seconds
Square 2
Hz
Triangle 2
Hz
Sine
0.05, 0.5, 1, 10, 25, 30, 40, 50, 60, and 100 Hz
R-Wave Detection
Width: 20 ms increments from 20 to 200 ms and
two smaller widths of 8 and 12 ms. Auto-step time
interval is 6 seconds.
Amplitude: 0.05 mV to 0.5 mV in 0.05 mV steps;
0.5 mV to 5.5 mV in 0.25 mV steps. Auto-step
time interval is 6 seconds.
R-Wave Rate: 60 BPM
Arrhythmia Selections
Supraventricular
Atrial fibrillation (1 and 2)
Atrial
flutter
Sinus
arrhythmia
Missed beat at 80 BPM
Missed beat at 120 BPM
Paroxysmal atrial tachycardia (PAT)
Nodal
rhythm
Supraventricular tachycardia (SVT)
Premature
Premature atrial contraction (PAC)
Premature nodal contraction (PNC)
Premature ventricular contraction (PVC)
Multifocal
1
Multifocal
2
Multifocal
3
Ventricular Rhythm
Bigeminy
Trigeminy
Couplet (pair of PVCs)
Run of 5 PVCs
Run of 11 PVCs