Function select - ecg - pacemaker menu – Fluke Biomedical medSim 300B User Manual
Page 51

Operation
Operating the medSim 300B
3
3-23
dem2
Demand 2. Mostly paced. Same as Demand 1 except 20 paced beats
and 40 normal beats.
avsq Atrial-ventricular
sequential.
Dual pacemaker spikes—first triggers
P-wave, second triggers QRST—0.15 seconds apart.
nonc
Non-capture. Asynchronous paced except every 10th beat there is no
QRST response.
nonf
Non-function. Pacemaker spikes only. No QRST response.
The ATRIAL PACER and the VENTRICLE PACER can be varied independently—the
amplitude (height) from -700 mV to +700 mV and the width from 0.1 ms to 2 ms.
To set atrial pacemaker parameters:
1. Press
MENU, and the following display:
ATRIAL PACER
WIDTH
AMP
eic049.eps
2. To set atrial pacemaker amplitude (height) and atrial pacemaker width
parameters, press the corresponding function keys, and then follow instructions
for step menu functions. The displays follow:
atrial height:
### mv
up
exc
auto
down
eic050.eps
atrial width:
#.## ms
up
exc
auto
down
eic051.eps