Equipment required for calibration, Calibration setup – Fluke Biomedical medSim 300B User Manual
Page 154

medSim 300B
Operators Manual
6-6
Equipment Required for Calibration
To begin the "on-site" calibration process, make sure you have the following equipment:
•
Digital multimeter- Keithley 196
*
Measurement Type
Measurement
Range
Accuracy (<±) Meter Output
Current
VDC
0-6 mV
10 µV
N/A
VDC
0-12 mV
20 µV
N/A
VDC
0-120 mV
200 µV
N/A
VDC
10 V
10 mV
N/A
Resistance 500
e
0.1
e
N/A
Resistance 15-25
ke 50
e
1-10
µV
Resistance 1-5
ke 5
e
10-100
µV
•
DC power supply - Instek Model PS-3225D
*
: 10 V ±10 mV @ 30 mA
•
Calibration cables
**
Part No.
Blood Pressure
2199081
Temperature
2199278
Cardiac Output
2199269
Note
*
Equivalent equipment is required to meet or exceed listed specifications.
Note
**
Cables are available from Fluke Biomedical. Refer to the current Fluke
Biomedical Price List for availability, part number information and price
Calibration Setup
The medSim 300B has no internal adjustments. All calibration adjustments are completed
using the medSim 300B keyboard. Calibration constants are stored in the nonvolatile
memory (EEPROM) within the medSim 300B.
1. Remove the Quality Seal from the bottom panel.
2. Then remove the three screws holding the metal plate and remove the metal
plate.
3. Locate the "CAL" switch near the rear of the opening.
4. Set the top panel switches as follow:
•
BASELINE IMPEDANCE
500
•
LEAD SELECT
II
•
SENSITIVITY 40
•
VAR TEMP-PROBE
700
•
CARDIAC OUTPUT/VAR
TEMP
VAR TEMP