Fluke Biomedical medSim 300B User Manual
Page 104

medSim 300B
Operators Manual
4-6
3. To zero all four of the medSim 300B blood pressure channels, press
from page 1 of the FUNCTION SELECT MENU, for the following two pages of
display:
BP CHANNEL SELECT
2
3
4
>
1
eic158.eps
BP CHANNEL SELECT
<
zero
CATH
eic159.eps
4. Press
This sets all the medSim 300B blood pressure channels to zero.
Note
Necessary to complete only once after connecting the medSim 300B to the
equipment under test (EUT).
5. Now zero the blood pressure channels to be used on the EUT.
To begin aortic valve simulation:
1. Choose
The medSim 300B display shows:
CATHETERIZATION
aort
mitr
pulm
eic160.eps
See also other documents in the category Fluke Biomedical Equipment:
- 7000DP Impulse Supplement (12 pages)
- 7000DP Impulse Getting Started (36 pages)
- 7000DP Impulse (92 pages)
- 06-526-2200 (26 pages)
- 07-417 (16 pages)
- 07-443 (14 pages)
- 07-451 (8 pages)
- 07-487 (12 pages)
- 07-453 (24 pages)
- 07-555 (8 pages)
- 07-553 (10 pages)
- 07-605-7777 (8 pages)
- 07-444 (46 pages)
- 07-618 (10 pages)
- 07-600 (6 pages)
- 07-591 (8 pages)
- SigmaPace 1000 (154 pages)
- 07-653 (8 pages)
- 07-633 (20 pages)
- 07-649 (14 pages)
- 07-661-7662 (12 pages)
- 07-645 (14 pages)
- 10100AT (80 pages)
- 07-644 (10 pages)
- 18-203 (4 pages)
- 07-621 (12 pages)
- 07-647 (12 pages)
- 18-207 (8 pages)
- 18-216-1000 (10 pages)
- 18-220 (10 pages)
- 18-228 (8 pages)
- 18-223 (18 pages)
- 18-229-1313 (6 pages)
- 18-250 (8 pages)
- 18-252 (6 pages)
- 18-303 (6 pages)
- 35035 (19 pages)
- 6000-528 (18 pages)
- 35080B (70 pages)
- 6000-529 (16 pages)
- 35080M (64 pages)
- 6000-530B (16 pages)
- 57-436 (12 pages)
- 57-440 (12 pages)
- 57-402 (32 pages)