Function select - cardiac output menu – Fluke Biomedical medSim 300B User Manual
Page 142

medSim 300B
Operators Manual
5-6
Function Select - Cardiac Output Menu
Set the "CARDIAC OUTPUT/VAR TEMP" switch on the top panel of the medSim 300B
to "CARDIAC OUTPUT".
Cardiac output simulation is set up:
•
By selecting an injectate temperature of 2 °C using the potentiometer on the CI-3
module/cable.
•
By selecting a baseline temperature from the medSim 300B CARDIAC
OUTPUT MENU.
•
By executing a waveform.
To begin the cardiac output simulation, adjust the 10-turn potentiometer on the CI-3
module/cable to simulate the 2 °C injectate temperature.
To set the baseline blood temperature from the medSim 300B:
1. Begin at the FUNCTION SELECT MENU and press the
access page 2 as shown below:
FUNCTION SELECT
<
C.O.
SEQ
DEFB
>
ARTF
eic216.eps
Note
Illustrations may not match your medSim 300B display if additions,
options, or expansion modules have been installed.
2. Press
CARDIAC OUTPUT
BASE
WAVE
TREND
ADJB
eic217.eps
3. Press
36˚
c.o. base temp
37˚
38˚
user
eic218.eps