Temperature section, Cardiac output section – Fluke Biomedical medSim 300B User Manual
Page 21

Chapter Title
medSim 300B Instrument Specifications
1
1-9
Channel 3 (mmHg):
Atmosphere (0)
Central venous pressure = 15/10
Right ventricle = 25/0
Pulmonary artery = 25/10
Pulmonary artery wedge = 10/2
Static: -10, -5, 0, 5, 10, 20 and 30
Manual or auto-stepping at 12 second intervals
Swan-Ganz: start, insert, inflate, deflate and remove
Channel 4 (mmHg):
Atmosphere (0)
Arterial = 120/80
Left ventricle = 120/0
Right ventricle = 25/0
Pulmonary artery = 25/10
Pulmonary artery wedge = 10/2
Triangle = 30, 2 Hz
Triangle = 300, 2 Hz
Static: -10, -5, 0, 20, 40, 80, 100, 200, 250 and 300
Manual or auto-stepping at 12 second intervals
Temperature Section
Variable Temperature:
34 °C, 37 °C, 40 °C, hyperthermia, hypothermia, and spike
Note
See the "Operation" chapter for descriptions of waveforms.
Note
"VAR TEMP" is inoperative during the Cardiac Output simulation.
Fixed Temperature:
Fixed at 37 °C
Channel 2's accuracy is 0.1 °C
Probe Compatibility:
400 and 700 series YSI types
Isolation:
Variable temperature is electrically isolated from the rest of the
instrument except Cardiac Output. Fixed temperature is isolated from
all other outputs.
Cardiac Output Section
See the "Cardiac Output" option chapter in this manual.