Artifact section, Ecg artifact, Blood pressure/respiration artifact – Fluke Biomedical medSim 300B User Manual
Page 22: Sequences section, Utility functions section, Rs-232 serial port interface

medSim 300B
Operators Manual
1-10
Artifact Section
ECG Artifact
Wave Type:
50 Hz, 60 Hz, muscle, or baseline-wander
Leads:
Any wave type can be added to any limb lead, or to the V-leads, or to all leads
Size:
0.25, 0.5, and 1.0 times the Lead II ECG amplitude setting
Blood Pressure/Respiration Artifact
Respiration artifact can be injected into any blood pressure waveform. Arterial and left
ventricle waveforms are modulated by the selected respiration rate at 5 % or 10 %. All
others have respiration added to them at 5 mmHg or 10 mmHg.
Sequences Section
Sequences step through a series of preprogrammed selections.
There are two types of sequences—autosequences and step-sequences. Autosequences
run automatically with no display interaction; step-sequences can be started and stopped
by the operator. There are three preprogrammed step-sequences and four
user-programmable step sequences.
Note
See the "Operation" chapter for more information and a list of the
sequences.
Utility Functions Section
View Angle:
Down, up and save; saved permanently
Auto Power Off Time:
0.5, 1.0, 2.0, 4.0 and 8.0 hours; and off.
This is reset each time a key is pressed; saved permanently
Key Beep:
On and off; not saved
RS-232 Baud Rate:
300, 600, 1200, and 2400 baud; saved permanently
Save:
Saves all current settings as power-on conditions
RS-232 Serial Port Interface
The medSim 300B can be interfaced to a personal computer or to the Fluke Biomedical
medTester 5000B. Most functions can be controlled remotely except for slide switch
controlled functions.
Note
A special cable is required and is available from Fluke Biomedical. Refer
to the current Fluke Biomedical Price List for availability, part number
information and price.
Baud rates are 300, 600, 1200, and 2400.