Fluke Biomedical medSim 300B User Manual
Page 71

Operation
Operating the medSim 300B
3
3-43
Selections Segments Run
Times
setup (Monitor setup)
Blood pressure/artifact off
ECG BPM 80
Respiration rate 20
BP1
atmosphere
BP2
atmosphere
BP3
atmosphere
BP4
atmosphere
Continuous
Stop
start (Monitor start)
ECG BPM 80
Respiration rate 20
BP1
artifact
BP2 pulmonary artery
BP3 central venous
pressure
BP3 left ventricle
Continuous
Stop
Function Select – Sequences - Step-sequences Menu
Step-sequences, as opposed to autosequences, can be started and stopped manually. To
select the STEP-SEQUENCES MENU, press
MENU:
STEP SEQUENCES
FIXD
USER
MAKE
REVU
eic106.eps
This menu provides:
•
Three preprogrammed (also referred to as "fixed" or "preset") step-sequences that
are accessed by pressing
•
Four user-programmable step-sequences that are accessed—only after being
programmed into the medSim 300B—by pressing
•
User step-sequences programming that is accessed by pressing
•
User step-sequences review that is accessed by pressing