Fluke Biomedical medSim 300B User Manual
Page 143

Cardiac Output
Cardiac Output Simulation
5
5-7
4. Press the corresponding function key for the desired baseline blood temperature.
Note
The default or normal baseline blood temperature selection is 37 °C.
Note
Changing the baseline blood temperature resets a running waveform.
Blood temperature selections and their descriptions follow:
Selections Descriptions
36 °
36 °C temperature
37 °
37 °C temperature
38 °
38 °C temperature
user
The temperature set from the ADJ BASE TEMP MENU
Note
ADJ BASE TEMP MENU instructions follow.
To set "user" baseline temperature:
1. Press
display:
adj base temp
down
up
eic219.eps
2. Press either
adjustment is reflected in the EUT display.
Note
This adjustment can be made after selecting any baseline temperature, but
the setting is lost on power-off except when saved from the UTILITY
MENU.
Note
To save a "user" baseline temperature setting, refer to the Function
Select - Utility Menu section in Chapter 3.