Fluke Biomedical medSim 300B User Manual
Page 115

Option 1
Cardiac Catheterization
4
4-17
Follow the illustration below.
Blood Pressure
Input
Cardiac
Catheterization
System
medSim 300B
Blood Pressure
Input
BP Cables
LV
BP3
PA OR PAW
Use BP1 and BP3 Blood Pressure
Channels for Mitral Valve Simulation
BP1
eic177.eps
Figure 4-13. Cardiac Catheterization Connections for Mitral Valve Simulation
1. Turn both units on.
2. After a connection has been made, the medSim 300B and the monitor need to be
zeroed.
3. To zero all four of the medSim 300B blood pressure channels, press
from page 1 of the FUNCTION SELECT MENU, for the following two pages of
display:
BP CHANNEL SELECT
1
2
3
4
>
eic178.eps
BP CHANNEL SELECT
<
zero
CATH
eic179.eps
4. Press
This sets all the medSim 300B blood pressure channels to zero.
Note
Necessary to complete only once after connecting the medSim 300B to the
equipment under test (EUT).
- 7000DP Impulse Supplement (12 pages)
- 7000DP Impulse Getting Started (36 pages)
- 7000DP Impulse (92 pages)
- 06-526-2200 (26 pages)
- 07-417 (16 pages)
- 07-443 (14 pages)
- 07-451 (8 pages)
- 07-487 (12 pages)
- 07-453 (24 pages)
- 07-555 (8 pages)
- 07-553 (10 pages)
- 07-605-7777 (8 pages)
- 07-444 (46 pages)
- 07-618 (10 pages)
- 07-600 (6 pages)
- 07-591 (8 pages)
- SigmaPace 1000 (154 pages)
- 07-653 (8 pages)
- 07-633 (20 pages)
- 07-649 (14 pages)
- 07-661-7662 (12 pages)
- 07-645 (14 pages)
- 10100AT (80 pages)
- 07-644 (10 pages)
- 18-203 (4 pages)
- 07-621 (12 pages)
- 07-647 (12 pages)
- 18-207 (8 pages)
- 18-216-1000 (10 pages)
- 18-220 (10 pages)
- 18-228 (8 pages)
- 18-223 (18 pages)
- 18-229-1313 (6 pages)
- 18-250 (8 pages)
- 18-252 (6 pages)
- 18-303 (6 pages)
- 35035 (19 pages)
- 6000-528 (18 pages)
- 35080B (70 pages)
- 6000-529 (16 pages)
- 35080M (64 pages)
- 6000-530B (16 pages)
- 57-436 (12 pages)
- 57-440 (12 pages)
- 57-402 (32 pages)