Cardiac output rs-232 serial port operation – Fluke Biomedical medSim 300B User Manual
Page 145

Cardiac Output
Cardiac Output RS-232 Serial Port Operation
5
5-9
2. Press the corresponding function key for the desired C.O. trend sequence.
Selections of C.O. trend sequences and listings of their segments follow:
Selections Descriptions
NORM
5.0, 4.5, 6.0, 5.5 and 4.0 L/min
INTR
4.5, 5.5, "intr", 5.5 and 4.5 L/min
SLOW
4.5, 5.5, "slow", 5.5 and 4.5 L/min
Upon pressing a function key related to one of the three C.O. trend sequences—NORM,
INTR, or SLOW—the top line of the display shows the segments of that particular trend
sequence.
For illustration purposes:
1. Press
5.0
strt
rset
4.5
6.0
5.5
4.0
eic223.eps
2. One of the segments (in this case, 5.0) has a cursor under it. Press
execute the marked segment. Upon execution, the marked segment (5.0) starts
flashing, and the cursor marks the next segment (4.5).
3. Press
then proceeds to mark the next segment (6.0). Continue in this manner to
complete the sequence.
Note
Flashing stops at the completion of a segment.
4. To reset the sequence, press
segment (5.0).
The three C.O. TRENDS MENU choices all follow the same pattern of menu
operation.
Cardiac Output RS-232 Serial Port Operation
By category, the following is a list of the medSim 300B Cardiac Output Option RS-232
remote commands and their descriptions.
Note
For complete information on remote commands, refer to the “RS-232 Serial
Port Operation” section in Chapter 3, and to the HHC3 Handheld
Controller Reference Manual..
Note
Each command is preceded by a bullet (•), which is not part of the
command string.