Late inflation, Early deflation – Fluke Biomedical medSim 300B User Manual
Page 130

medSim 300B
Operators Manual
4-32
Late Inflation
Dicrotic Notch
Inflation
Deflation
Assist Ratio = 1:2
Unassisted
Waveform
Assisted
Waveform
eic207.eps
Figure 4-27. medSim 300B Simulated Late Balloon Inflation
•
Inflation begins after the dicrotic notch (aortic valve closure).
•
Deflation ends just before the next arterial pressure waveform.
Note
The dicrotic notch is visible prior to the balloon inflation point.
Early Deflation
Dicrotic Notch
Inflation
Deflation
U Wave
Assist Ratio = 1:2
Unassisted
Waveform
Assisted
Waveform
eic208.eps
Figure 4-28. medSim 30013 Simulated Early Balloon Deflation
•
Inflation begins at the dicrotic notch (aortic valve closure).
•
Deflation ends too soon, causing a large U-wave before the next arterial
waveform.
See also other documents in the category Fluke Biomedical Equipment:
- 7000DP Impulse Supplement (12 pages)
- 7000DP Impulse Getting Started (36 pages)
- 7000DP Impulse (92 pages)
- 06-526-2200 (26 pages)
- 07-417 (16 pages)
- 07-443 (14 pages)
- 07-451 (8 pages)
- 07-487 (12 pages)
- 07-453 (24 pages)
- 07-555 (8 pages)
- 07-553 (10 pages)
- 07-605-7777 (8 pages)
- 07-444 (46 pages)
- 07-618 (10 pages)
- 07-600 (6 pages)
- 07-591 (8 pages)
- SigmaPace 1000 (154 pages)
- 07-653 (8 pages)
- 07-633 (20 pages)
- 07-649 (14 pages)
- 07-661-7662 (12 pages)
- 07-645 (14 pages)
- 10100AT (80 pages)
- 07-644 (10 pages)
- 18-203 (4 pages)
- 07-621 (12 pages)
- 07-647 (12 pages)
- 18-207 (8 pages)
- 18-216-1000 (10 pages)
- 18-220 (10 pages)
- 18-228 (8 pages)
- 18-223 (18 pages)
- 18-229-1313 (6 pages)
- 18-250 (8 pages)
- 18-252 (6 pages)
- 18-303 (6 pages)
- 35035 (19 pages)
- 6000-528 (18 pages)
- 35080B (70 pages)
- 6000-529 (16 pages)
- 35080M (64 pages)
- 6000-530B (16 pages)
- 57-436 (12 pages)
- 57-440 (12 pages)
- 57-402 (32 pages)