In the step menu, the top display line indicates the name of the parameter being

modified and a value for it. If the value shown (illustrated above as "#.##") is

active, it flashes.

2. Press -DOWN or -UP to decrement or increment the displayed value.
3. Press -EXC to execute the displayed value.
4. Press -AUTO and the bottom display line changes as follows:

ecg amp, lead II: #.## mv

down

up

strt

man

eic005.eps

In this mode, use of the -DOWN key and the -UP key decrements and

increments the displayed value, and immediately executes the selected value

without requiring a separate "execute" key.

Note

Use of the -MAN key returns to the previous mode.

5. To activate the automatic stepping mode, press the -STRT key and the

display changes to:

ecg amp, lead II: #.## mv

down

up

stop

eic006.eps

At this point, the parameter is automatically stepping at a predetermined rate

(refer to the "Specifications" chapter). Upon reaching either limit, the stepping

switches directions and continues.

6. Choose -DOWN or -UP to reverse the incrementing and decrementing

of the values.

7. Press -STOP to stop the stepping and return to the manual mode.
8. Pressing the also stops automatic stepping and moves you back up

through the menu.
Only one automatic stepping mode can be active at a time.