Fluke Biomedical medSim 300B User Manual
Page 81

Operation
Operating the medSim 300B
3
3-53
Function Select - Defibrillator Tests -Emergency 2 Menu
"Emergency 2" selects a series with 10 normal beats, 25 beats of ventricular tachycardia
(180 BPM), and then a continuous fine ventricular fibrillation.
To access the EMERGENCY 2 MENU:
1. Press
emergency
2:
start
hold
eic127.eps
Menu selections (same as "EM1") and their descriptions follow:
Selections Descriptions
start
Starts the series
hold
Disables the signal from the "SYNC/A PACE" input. When "hold" is
flashing, the signal is disabled
2. Press
emergency
2: running
start
hold
eic128.eps
When the waveform changes to ventricular fibrillation, the display changes to:
emergency
2: vfib
start
hold
defib
eic129.eps
3. Press
PACE" jack to defibrillate the simulator and convert the ECG to normal sinus
rhythm.
Note
For the medSim 300B patient simulator to be able to receive the input
signal, make certain that "hold" is not flashing.
A successful defibrillation is indicated as the display changes to: