To select BP 4 static pressures, press
MENU for the following display:
bp 4 static:
## mmHg
down
up
exc
auto
eic092.eps
Follow the instructions for the step menu functions as previously described in this
chapter.
Function Select - Temperature Menu
To access the TEMPERATURE MENU, from the FUNCTION SELECT MENU:
1. Press
Note
Temperature selections are available only when the "CARDIAC
OUTPUT/VAR TEMP" switch is set to "VAR TEMP". For more
information, refer to the Temperature—Variable and Fixed section in the
PANEL DESCRIPTIONS, CONTROLS, AND CONNECTORS section near
the beginning of this chapter.
TEMPERATURE MENU selections are shown in two pages of display as
follows:
temperature
34º
37º
40º
hypr
>
eic093.eps
temperature
hypo
spike
<
eic094.eps
2. Press the corresponding function key for the desired temperature selection.
Temperature selections and their descriptions follow:
Selections Descriptions
34 °
34 °C fixed temperature.
37 °
37 °C fixed temperature.