Accessories, Standard, Optional – Fluke Biomedical medSim 300B User Manual
Page 140: Cardiac output simulation, Introduction

medSim 300B
Operators Manual
5-4
Trends: Normal
sequence
Sequence with interrupted injectate
Sequence with slow injectate
Accuracy:
10 % of setting for 3 L/min, 5 L/min, and 7 L/min at 37 °C
±0.4 °C for baseline temperature
Isolation:
Cardiac output is electrically isolated from the rest of the
unit except temperature channel 1
Note
Cardiac output is inoperative during the temperature channel 1 simulation
("VAR TEMP" mode).
Accessories
Standard
Accessories Part
No.
CI-3 module/cable
2392199
General purpose connector
2392158
Note
Additional general purpose connectors are available from Fluke
Biomedical.
Optional
Currently offered cardiac output adapter cables are listed below:
•
Hewlett Packard Cardiac Output Adapter
•
Hewlett Packard Adapter
•
Gould/SpectraMed Model 1445 Adapter
•
Gould/SpectraMed Model 1465 Adapter
Note
Refer to the current Fluke Biomedical Price List for availability, part
number information, and price.
Cardiac Output Simulation
Introduction
The Cardiac Output Option for the medSim 300B consists of both internal firmware and
external hardware.
The external hardware is designated as the CI-3 module/cable, and plugs directly into the
medSim 300B top panel at the "CARDIAC OUTPUT" jack.
Note
See illustration "CI-3 module/cable" located at the end of this chapter.