Blood pressure section – Fluke Biomedical medSim 300B User Manual
Page 20

medSim 300B
Operators Manual
1-8
either "positive" or "negative". The rate is shifted to
120 BrPM for 12 seconds each minute.
Ventilator Simulation:
40 BrPM at fixed ratio, other parameters variable as in
"normal"
Blood Pressure Section
Input/Output Impedance:
300 e
Exciter Input Voltage Range:
2 to 16 V
Exciter Input Frequency Range:
DC to 4 kHz
Output Sensitivity:
5 or 40
μ
V/V/mmHg
Output Range:
-10 to +300 mmHg on BP1, BP2 and BP4
-10 to +30 mmHg on BP3
Accuracy:
±
(1 % of full range + 1 mmHg) at 80 BPM, normal sinus
rhythm ECG only
Rate:
All dynamic pressures track all normal sinus rhythm rates
and track all arrhythmias
Isolation:
Blood pressure circuitry is electrically isolated from all
other medSim 300B outputs
Channel 1 (mmHg):
Atmosphere (0)
Arterial = 120/80
Left ventricle = 120/0
Central venous pressure = 15/10
Right ventricle = 25/0
Pulmonary artery = 25/10
Pulmonary artery wedge = 10/2
Static: -10, -5, 0, 20, 40, 80, 100, 200, 250, and 300
Manual or auto-stepping at 12 second intervals
Channel 2 (mmHg):
Atmosphere (0)
Arterial = 120/80
Left ventricle = 120/0
Central venous pressure = 15/10
Right ventricle = 25/0
Pulmonary artery = 25/10
Pulmonary artery wedge = 10/2
Static: -10, -5, 0, 20, 40, 80, 100, 200, 250 and 300
Manual or auto-stepping at 12 second intervals