Calibration procedure – Fluke Biomedical medSim 300B User Manual
Page 155

Performance Checks/Calibration
Calibration
6
6-7
5. Plug the medSim 300B into the battery eliminator. Press and hold down the
"CAL" switch.
6. Turn on the medSim 300B and continue to hold down the "CAL" switch until the
display reads:
TEST FUNCTION SELECT
HDWR
RCAL
CAL
EXIT
eic257.eps
7. Press
Calibration Procedure
During calibration, the top line of the display shows the name of the calibration step,
where to connect the digital multimeter (DMM) and the value to adjust for.
Some of the steps require only that a DMM reading be taken before proceeding to the
next step.
For steps that require adjustment, the menu selections are displayed on four menu pages,
which are accessed using the
that relates to the desired selection.
To decrement and increment, use the
the setting and jumps back a step.
Menu selections and their functions follow:
Selections Functions
sdn slow
decrement
sup slow
increment
mdn medium
decrement
mup medium
increment
fdn fast
decrement
fup fast
increment
next
moves to the next step
skip
moves to the next step without saving the new setting
esc
exits the calibration procedure without saving the new setting
save
exits the calibration procedure and saves the new setting
1. For the "ECG: REF" step, record the reading (including polarity).
2. Then on the next step—"ECG: 5 mV"—adjust the unit for a reading of 5 mV
plus the reference step reading. This can also be done by zeroing the DMM on
the reference step.
3. For "PACE: REF", record the reading as above. Then for the "ATRIAL PACE:
REF V" and "VENT PACE: REF V", adjust the unit to the "PACE REF" voltage.